Abigail Koss, TOFWERK USA
Ethylene oxide is an emerging contaminant of increasing interest to regulatory agencies. Commonly used, real-time, direct mass-spectrometric methods – specifically, proton-transfer-reaction (PTR) – face severe limitations in monitoring ethylene oxide in ambient air due to interferences. Ethylene oxide with typical background concentration of hundreds of part-per-trillion (ppt) shares the same protonated mass-to-charge ratio as acetaldehyde (45 Th, C2H4O•H+), which is ubiquitous in the atmosphere at typically much higher background concentrations.
Ethylene Oxide Detection with the Vocus CI-TOF
The following method using the Vocus CI-TOF mass spectrometer equipped with the Vocus PTR Reactor allows sensitive, real-time measurement of ethylene oxide in ambient air without interference from acetaldehyde. This method can be additionally used for the detection of aromatics (e.g. BTX) and terpenes. It uses ultra-high purity (UHP) oxygen (O2) as a reagent gas to generate O2+ reagent ions. The target analytes can undergo ionization through three distinct pathways including charge transfer for aromatic compounds (BTX), hydride abstraction for aldehydes (propanal, acetaldehyde), and adduct formation for ethylene oxide.
Results
In Figure 1 two overlaying mass spectra of two separate measurements of ethylene oxide and acetaldehyde, both at mixing ratio of 20 parts per billion (ppb), are presented. The figure illustrates the distinct detection of ethylene oxide as a cluster C2H4O•O2+ at mass-to-charge ratio 76. Additionally, it highlights the minimal interference from acetaldehyde, which is less than 0.5%. Acetaldehyde can be detected as C2H3O+ at mass-to-charge ratio 43 Th.
In Figure 2 a six-point calibration curve relating the concentration of ethylene oxide to the measured signal is shown. The curve demonstrates that the analytical method exhibits a linear response while the slope indicates the sensitivity which is equal to 59 counts per second (cps) per ppb.
In Table 1, calculated limits of detection (LODs), defined as three times standard deviation of signal in clean air, for ethylene oxide and other selected species are reported. The LODs were directly measured using a Vocus 2R model. The LODs for other models present in the table were derived using typical relative sensitivities for such models.
Compound | Detected as | Model 2R/S | Model Scout | Model Eiger | |||
1-sec. | 1-min. | 1-sec. | 1-min. | 1-sec. | 1-min. | ||
ppt | ppt | ppt | |||||
Ethylene oxide | C2H4O•O2+ | 270 | 29 | 400 | 40 | 850 | 90 |
Benzene | C6H6+ | 14 | 1.0 | 20 | 2 | 40 | 3 |
Toluene | C7H8+ | 11 | 0.80 | 15 | 1 | 40 | 3 |
Xylene | C8H10+ | 15 | 1.2 | 20 | 2 | 40 | 4 |
Trimethylbenzene | C9H12+ | 19 | 1.9 | 30 | 3 | 60 | 6 |
Propanal | C3H5O+ | 150 | 13 | 250 | 20 | 500 | 40 |
α-pinene | C10H16+ | 100 | 11 | 150 | 15 | 350 | 30 |
Figure 3 illustrates real-time simultaneous measurement of acetaldehyde and ethylene oxide in ambient urban air. As anticipated, acetaldehyde exhibits elevated mixing ratios and frequent spike events in the measurements, reflecting its relatively higher prevalence in the atmosphere. On contrary, ethylene oxide, which is scarcer on atmospheric sources, maintains a steady background level. The distinct time evolution provides further evidence of the possibility of differentiation these two species within one measured dataset.
Summary
The analytical method described in this application note can be valuable in various scenarios. Real-time measurement of ethylene oxide is essential for ensuring workplace safety in industrial settings, in environmental stationary or mobile monitoring evaluating potential contamination or compliance with air standards. Further, this technique can be used in emergency response, providing rapid, on-site detection of ethylene oxide during chemical spills. It is also well suited for product quality control where headspace sample analysis serves as verification that the product underwent ethylene oxide treatment such as for medical devices, pharmaceuticals, or personal care products.
Recommended Vocus Models for Ethylene Oxide Detection
This method is compatible with Vocus CI-TOF Eiger, Scout, S, and 2R models, when utilizing the Vocus PTR Reactor. Instruments already equipped with selective reagent ion option (O2+ and/or NO+) can use this method without requiring any hardware alterations. Instruments only capable of H3O+ ion chemistry may use this method with minor hardware modifications. The only additional consumable required is UHP oxygen. The Vocus Elf model is not recommended for this measurement due to its unit mass resolving power which does not permit the separation of ethylene oxide from an isobaric interference in ambient air. This model could be used in scenarios where the isobaric interference is insignificant compared to ethylene oxide.